Inspection Function on Rear Panel (for DM3054 ONLY)

CAUTION:

Voltage between LO terminal and ground limit to 150Vpeak(Max).

Channels: 12 differential voltage channels, 4 differential current channels.
Measurement Functions: 2WR, Capacitance, DCV, DCI, ACV, ACI, Diodes , frequency and Period.

Work characteristic: Thermo EMF <6uV. Maximum scanning rate is 2 channels per second.

Input characteristic: Differential input voltage 150Vpeak(Max), isolation voltage between channels

150Vpeak(Max), current input terminal 1Apeak(Max), channel isolation >60dB(@10kHz), all
terminal to the chassis ground voltage 150Vpeak(Max).

Current channel protection: Internal 2A self- recover fuse
Voltage channel protection: 250V over voltage protection.


Real-time Clock
Precision:

1min/month (Environment Temperature >0℃)

Clock battery Life:

2 years


Math Functions
Null, Min/Max/Average, dBm, dB, Limit Test (with TTL output)

Other Functions
Reading Hold, Ratio Measurement

High-speed Sampling
50kSa/s (In Datalog function)

Reading Resolution
480,000 Count, >5 3/4 digits

USB I/O Interface
USB Host (support U-disk), USB Device interface.

Other I/O Interface
RS232, GPIB (Optional) support for SCPI command, LAN (Optional)