RIGOL M300 User Manual

Page 63

Advertising
background image

RIGOL

© 2007 RIGOL Technologies, Inc.

User’s Guide for DM3000 Series

2-15

Limit Measurement

Use the Limit test function to perform pass/fail testing with respect to specified
upper and lower limits. The limits can be set to any value between 0 and ± 120% of
the highest range of the current function. The upper limit must be positive than the
lower limit.

Press

 Limit, the display shows:

Figure 2- 12 Limit Setting

Table 2- 11 Limit Measurement Menu Function Description

Table 2- 12 Limit Parameters

Measurement

Description

DC Voltage

-1200V ~ +1200V

AC Voltage

0~ 900V

DC Current

-12A ~ +12A

AC Current

0~ 12A

Resistance

0 ~ 120MΩ

Capacitance

0 ~ 240uF

Frequency

3Hz ~ 300kHz

Period

3us ~ 300ms

Ratio

-1G ~ 1G


The range of Limit function:
(1). The upper limit value should be always greater than the lower limit value.
(2). The upper/lower limit values are stored in volatile memory.

Function

Menu

Settings

Description

High

Set the desired Upper limit.

Low

Set the desired Lower limit.

Save all changes, back to a higher level menu.