Period Test

Figure 1- 34 Period Measurement Data Interface

Table1- 11 Period Test Characteristics

Range

200mV, 2V, 20V, 200V, 750V

Measurement Range

0.33s ~ 3.3us

Input Signal Range

100mVAC~750VAC

Import Protection

750V

RMS

on all ranges (HI

Terminal)

Configurable Parameters Null value


Steps:
1. Connect test leads as Figure 1- 35 shown. Red test lead connects the HI

Terminal, Black test lead connects the LO Terminal.

2. Press

twice to select the Period Test.

3. Set the Null value.

Null computing will be an optional operation, could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.

(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set up Measurement Parameters”, Null computing)