Wire resistance measurement – RIGOL M300 User Manual

Page 27

Advertising
background image

RIGOL

© 2007 RIGOL Technologies, Inc.

User’s Guide for DM3000 Series

1-17

4-Wire Resistance Measurement

Figure 1- 21 4-Wire Resistance Measurement Interface

Table1- 6 Resistance Measurement Characteristics

Seven Ranges

200Ω, 2kΩ, 20kΩ, 200kΩ, 1MΩ, 10MΩ, 100MΩ

Max Resolution 100uΩ

Open-circuit

Voltage

<7V

Import Protection

(1). 200V

PK

(2). 1000V on all ranges (HI

Terminal)

(3). 200V on all ranges (HI Sense, LO Sense)

Configurable

Parameters

Range, Null value


Steps:
1. Connect test leads as show in Figure 1- 22; red test lead to the HI Terminal,

black test lead to the LO Terminal.

2. Press

twice to select the 4-Wire Resistance Measurement.

3. Choose the appropriate measurement range.
4. Set the Null setting value.

Null computing will be an optional operation, it could be setup in accordance
with users’ demand. If user does not implement Null computing, this parameter
is not required, direct implementation of the next step.

(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set up Measurement Parameters”, Null computing)
5. Lead test leads into circuit, start to measure.