dBm Measurement

This function applies to AC voltage and DC voltage measurements only.
The dBm function is logarithmic, and is based on a calculation of power delivered to
a reference resistance, relative to 1 milliwatt.

Press  dBm, the display shows:

Figure 2- 14 dBm Measurement


Table 2- 14 dB Measurement Function Menu Function Description

Function

Menu

Description

Default

Use the default value.

Save all changes, back to a higher level menu.



The computation method of the dBm:
dBm = 10 x Log

10

[ (Reading

2

/ R

REF

) / 0.001W ]

R

REF

expressed measuring the resistance value in the actual electric circuit.