To measure ac current, To measure ac current -13 – RIGOL M300 User Manual

Page 23

Advertising
background image

RIGOL

© 2007 RIGOL Technologies, Inc.

User’s Guide for DM3000 Series

1-13

To Measure AC Current

The following shows the system connections and selection of measurement functions.
The practice provides as guide to be familiar with the AC Current measurement
technique. (The AC functions only support 5½ digits measurement.)

Figure 1- 15 AC Current Measurement Data Interface

Table1- 4 AC Current Measurement Characteristics

Five Ranges

20mA, 200mA, 1A, 10A

Max Resolution 100nA

Input Protection

10A, 250V Current Input Fuse on rear
panel

Configurable

Parameters

Range, AC Filter, Null value


Steps:
1. Connect test leads as shown in Figure 1- 16; red test lead to the I Terminal,

black test lead to LO Terminal.

2. Press

to select the AC Current measurement function.

3. Choose the appropriate measurement range.
4. Setup the AC Filter.

Press

 Filter, to setup the AC Filter Bandwidth (Default value: “Mid”

(Middle)).

5. Set the Null setting value.

Null computing will be an option operation, could be setup in accordance with

Advertising