To measure ac voltage, To measure ac voltage -9 – RIGOL M300 User Manual

Page 19

Advertising
background image

RIGOL

© 2007 RIGOL Technologies, Inc.

User’s Guide for DM3000 Series

1-9

To Measure AC Voltage

The following shows the system connections and selection of measurement functions.
This practice provides a guide to get familiar with the AC Voltage measurement
technique. (The AC functions only support 5½ digits measurement.)

Figure 1- 9 AC Voltage Measurement Data Interface

Table1- 2 AC Voltage Measurement Characteristics

Five Ranges

200mV, 2V, 20V, 200V, 750V

Max Resolution 100nV

Input Protection 750V

RMS

on all ranges (HI

Terminal)

Configurable

Parameters

Range, AC Filter, Null value


Steps:
1. Connect test leads as shown in Figure 1- 10; red test lead to the HI Terminal,

black test lead to the LO Terminal.

2. Press

to select the AC Voltage measurement function.

3. Choose the appropriate measurement range.
4. Setup the AC Filter.

Press

 Filter, to setup the AC Filter Bandwidth (Default value: Mid).

5. Set the Null value.

Null computing will be an option operation, could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.