Ac filter – RIGOL M300 User Manual

Page 52

Advertising
background image

RIGOL

© 2007 RIGOL Technologies, Inc.

User’s Guide for DM3000 Series

2-4

AC Filter

There are three settings at the AC Filter menu. Choose the appropriate setting for
more accurate measurements. This function applies AC Voltage and AC Current
measurement only.

Press

 Filter, enter the menu shown below:

Figure 2- 2 Set the AC Filter

Table 2- 2 Menu Description

Function

Menu

Description

Slow

Set up the filter with low speed.

Mid

Set up the filter with to middle speed.

Fast

Set up the filter with high speed.

Save all changes, back to a higher level menu.

Table 2- 3 AC Filter Parameters Characteristics

AC Filter Options

Input Frequency

Setting Timer

Slow

3Hz~300kHz

1.2 reading/s

Mid

20Hz~300kHz

0.5 reading/s

Fast

200Hz~300kHz

0.3 reading/s

AC Filter
The AC Filter Parameters are stored in nonvolatile memory. The default is “Mid”
(middle).