Digital Communications Interface

Commands

6

6-43

Caution

If the password is forgotten, the themo-hygrometer must be

returned to an Authorized Service Center to have the password

restored.

SYSTem:PASSword:RECord?

Returns the state of data recording settings password-protection.
Example: SYST:PASS:REC?
Response: 0
A value of 1 is returned if data recording password settings are password-protected;

otherwise a value of 0 is returned.

SYSTem:PASSword:RECord <bool>

Sets the state of data recording settings password-protection.
Example: SYST:PASS:REC 1
The <bool> parameter turns the data recording password-protecton on (1 or ON) or off (0

or OFF). This command is password protected. Commands protected with this setting

are: DATA:REC:CLE, DATA:REC:FEED:RHUM, DATA:REC:FEED:TEMP, and

DATA:REC:TIM.

SYSTem:PASSword:SENSor?

Returns the state of the sensor settings password-protection.
Example: SYST:PASS:SENS?
Response: 0
A value of 1 is returned if the sensor settings password-protection is enabled. Otherwise,

a value of 0 is returned.

SYSTem:PASSword:SENSor <bool>

Sets the state of the sensor settings password-protection.
Example: SYST:PASS:SENS 1
The <bool> parameter turns the sensor settings password-protection on (1 or ON) or off

(0 or OFF). This command is password protected. Commands protected with this setting

are: CALC:PAR:DATE:TIME, CALC:PAR:RES, ROUT:CLOS, ROUT:OPEN,

SENS:AVER, SENS:IDEN, SENS:LOCK,and TRIG:TIME.

SYSTem:PASSword:STATistics?

Returns the state of the statistics settings password-protection.
Example: SYST:PASS:STAT?
Response: 0
A value of 1 is returned if the statistics settings password-protection is enabled,

otherwise, a value of 0 is returned.

SYSTem:PASSword:STATistics <bool>

Sets the state of the statistics settings password-protection.