Maintenance, Troubleshooting, Incorrect temperature or humidity reading – Fluke 5020A User Manual

Page 133: Missing or scattered graph data

Advertising
background image

Maintenance and Troubleshooting

Maintenance

8

8-3

Maintenance

The calibration instrument has been designed ease of operation and simplicity of

maintenance as the central theme in the product development. Therefore, with proper care

the instrument should require very little maintenance. Avoid operating the instrument in

oily, wet, dirty, or dusty environments.

If the outside of the instrument becomes soiled, it may be wiped clean with a

damp cloth and mild detergent. Do not use harsh chemicals on the surface, which

may damage the LCD or the plastic front panel.

If a hazardous material is spilt on or inside the equipment, the user is responsible

for taking the appropriate decontamination steps as outlined by the national

safety council with respect to the material.

Before using any cleaning or decontamination method except those

recommended by Fluke, users should check with an Authorized Fluke Scientific

Service Center to be sure that the proposed method will not damage the

equipment.

If the AC adapter becomes damaged, have it replaced immediately. Never

disassemble the AC adapter or attempt to repair it.

If the instrument is used in a manner not in accordance with the equipment

design, the operation of the thermo-hygrometer may be impaired or safety

hazards may arise.

Periodic calibration of the sensors at regular intervals is required to ensure that it

continues to measure with proper accuracy. Calibration should only be performed

by qualified authorized personnel.

Troubleshooting

In case you run into difficulty while operating the thermo-hygrometer, this section

provides some suggestions that may help you solve the problem. Below are several

situations that may arise followed by possible causes and suggested actions you might

take.
The following sections list possible situations, their causes, and suggested actions.

Incorrect temperature or humidity reading

While attempting to measure temperature or humidity, the display shows an incorrect

value or no value at all ( . . . . . . . . . ).

Check to see if the sensor is properly seated in the correct sockets.

If the actual temperature is too low or too high there may be a problem with the

sensor.

The measure period may be set to a long interval. Check the measure period

setting.

Refer to Sensors for information on proper sensor handling and situations that

can affect accuracy.

Missing or Scattered Graph Data

If a graph is empty or consists of only scattered points, it may be due to data recording

not being enabled or the recording period set too large. See Record Setting.

Advertising