Graph scale – Fluke 5020A User Manual

Page 47

Advertising
background image

Menu Functions

Display Menu

5

5-11

avu013.jpg

Figure 5-12. Field Data

Use the B C buttons to change each of the settings. Press E to save and continue to the

next parameter. Press the F button to cancel any changes and exit to the menu.

Graph Scale

The GRAPH SCALE function allows the selection of the graph settings for each zone (1,

2, 3, or 4). Graph settings for each zone are used in common among all display layouts.

This may be password-protected. For each zone you may set the center of the vertical

axes for the temperature graphs (T CENT), the temperature scale for the axes (T SCAL),

the center of the vertical axes for the humidity graphs (H CENT), the humidity scale for

the axes (H SCAL), the time axis scale (time), and the number of grid divisions for the

graph (DIVS). Following is the range for each setting:
T CENT from 100.000 to -40.000
T SCAL from 100.000 to 0.100
H CENT from 100.000 to 0.000
H SCAL from 100.000 to 1.000
TIME from 1, 4, 10, 24 hours and 2, 5, 7, 15, 30 days
DIVS from 1, 2, 4, 6, 8, 10