Calibration Procedure

Introduction

7

7-3

Introduction

5026A-x sensors used with the thermo-hygrometer should be periodically calibrated to

ensure accuracy. A six-month calibration interval for the first year, and one-year

thereafter, is recommended. The thermo-hygrometer display unit does not require

calibration, and the sensor can be sent to a qualified service center for calibration without

the thermo-hygrometer display unit. Contact a Fluke Authorized Service Center for return

instructions prior to returning the probe for calibration. Ensure that the probe is packaged

adequately to ensure safe transport. It is recommended that you acquire and use the

optional probe case for transport.
For those who wish to perform their own calibration, the calibration procedure is outlined

as follows.

Equipment

Temperature/humidity chamber to use as a relative humidity reference and a stable-

temperature environment (0.37% RH uncertainty or better is recommended)

PRT and associated readout to use as the reference thermometer (0.03 °C uncertainty

or better is recommended)

5020A Thermo-Hygrometer to use to read the 5026A-x sensor

An extension cable to connect with the 5026A-x sensor placed inside the

temperature/humidity chamber

Preparation

Allow the 5026A-x sensor to “dry out” in air with relative humidity between 20 and 50%

and temperature between 16 and 30 °C for 24 hours before calibration. Place the 5026A-x

sensor inside the temperature/humidity chamber, near the center, attached to an extension

cable that runs to the 5020A outside the chamber. Place the PRT inside the chamber with

the tip close to the tip of the 5026A-x sensor. Run the probe cable to the outside of the

chamber and connect it to the PRT readout.

Calibration Points

Each measurement parameter, both temperature and relative humidity, is calibrated at

three points throughout its range with the other parameter held constant at the center of

its range. The following tables list the calibration points for the two sensor models and

for both measurement parameters.

Table 7-1. 5026A-S Temperature Calibration Points

Temperature

At Relative Humidity

15 °C

45%

25 °C

45%

35 °C

45%

Table 7-2. 5026A-S Relative Humidity Calibration Points

Relative Humidity

at Temperature