Interface commands, Command summary, Command syntax – Fluke 5020A User Manual

Page 74

Advertising
background image

5020A

Users Manual

6-4

Interface Commands

Command Summary

An alphabetical listing of the commands implemented by the thermo-hygrometer readout

are shown in Table 6-1. These commands can be used with either serial or IR

communications ports.

Command Syntax

The thermo-hygrometer readout accepts commands that set parameters, execute functions

or respond with requested data. These commands are in the form of strings of ASCII-

encoded characters. As far as possible, the thermo-hygrometer readout conforms to SCPI-

1994. One notable exception is that compound commands are not allowed as explained

below.
Commands consist of a command header and, if necessary, parameter data. All

commands must be terminated with either a carriage return (ASCII 0D hex, 13 decimal)

or new line character (ASCII 0A hex, 10 decimal).
Command headers consist of one or more mnemonics separated by colons (:).

Mnemonics may use letter characters, the underscore character (_), and possibly numeric

digits as well. Commands are not case sensitive. Mnemonics often have alternate forms.

Most mnemonics have a long form that is more readable and a short form consisting of

three or four characters that is more efficient.
A mnemonic may end with a numeric suffix that specifies one of a set of independent

function blocks such as input channel data paths. If a numeric suffix is omitted when a

particular block must be specified, an error is generated (“Header suffix out of range”).
Query commands are commands that request data in response. Query commands have a

question mark (?) immediately following the command header. Responses to query

commands are generated immediately and placed in the output buffer. Responses are then

transmitted automatically over the RS-232 or IR ports. Responses are lost if not read

before the next command is received.
Some commands require parameter data to specify values for one or more parameters.

The command header is separated from the parameter data by a space (ASCII 20 hex, 32

decimal). Multiple parameters are separated by a comma (,).
The thermo-hygrometer readout does not allow compound commands (multiple

commands per line separated with semicolons). All commands are sequential. The

execution of each command is completed before subsequent commands are processed.

Table 6-1. Interface Commands

Command Description

*CLS

Clear the status registers

*ESE?

Returns the Standard Event Status Enable register

*ESE <num>|MIN|MAX|DEF

Sets the Standard Event Status Enable register

*ESR?

Returns the Standard Event Status register

*IDN?

Returns the instrument identification string that
indicates the manufacturer, model number, serial
number, and firmware version