Introduction, Dc power source, Battery – Fluke 5020A User Manual

Page 33

Advertising
background image

General Operation

Introduction

4

4-3

Introduction

This section explains basic operation of the thermo-hygrometer. Detailed operation of the

thermo-hygrometer is explained in Chapters 5 and 6. Chapter 5 explains the menu

structure and the functions available in the menu structure and Chapter 6 explains the

communications interface for operating the thermo-hygrometer remotely.

DC Power Source

The thermo-hygrometer requires 12V DC to operate. The AC adapter is provided to

produce the DC power from an AC mains supply.

Caution

For CE compliance and for proper performance, use only the

AC adapter shipped with the instrument. If the AC adapter

needs to be replaced, contact an Authorized Service Center.

The AC adapter has circuits with high voltages inside that could

present danger of electric shock or fire if exposed. If the AC

adapter is damaged in any way or becomes hot, discontinue

use immediately, disconnect the adapter from any AC supply,

and replace the adapter. Do not attempt to open, repair, or

continue using a damaged or defective AC adapter.

The DC output from the AC adapter plugs into the 12 V DC power input on the left side

of the instrument (see Figure 3-3).

Battery

The thermo-hygrometer uses a 9V battery to maintain continuous measurement and

recording during power outages. The recommended battery type is a standard 9V alkaline

battery (NEDA 1604A or IEC 6LR61). During a power outage, or when the AC adapter

is disconnected, the display will be inoperable but measuring will continue if the 9V

battery is installed. If the alarm is enabled, the beeper will sound periodically to alert the

user of the loss of external power. With a fresh alkaline battery, measuring will continue

during a power outage for typically about 16 hours. During normal operation, the battery

charge is regularly checked, and if the low battery alarm is enabled, the user will be

alerted when the battery charge drops below approximately 50%. The battery charge

level can be viewed with a display layout that includes a stat type zone with one of the

fields set to BATT. To prevent inadvertent battery drain, remember to switch the power

switch off when external power is disconnected and the thermo-hygrometer is not in use.
To install or replace the battery, follow these steps:

1. Switch the power off and unplug the DC power cord.
2. Turn the thermo-hygrometer over to access the rear battery compartment. Press

down slightly on the battery cover and slide it out to remove it.

3. Remove the old battery if one is in place by lifting it up at the bottom and pulling

it out.

4. Slide the new battery in at an angle, observing the proper polarity so the

terminals mate correctly, then press the bottom of the battery down into the

holder.

5. Replace the battery cover.
6. Replace the DC power cord and switch the power on.

Advertising