Figure 5-3. Channel Menu

Channel Setting

The CHANNEL SETTING function allows you to turn the channels on or off, set the

measurement period, turn averaging on or off, and set the time over which the rate of

change is calculated for measurements.
These settings may be password-protected to prevent them from being changed. The

channels are turned on or off by selecting the channel and then the ON or OFF setting

using the B C buttons.
MEA PER, measurement period, is the time interval at which the measurements are

updated on the display. You may select between 1, 2, 5, 10, 15, and 30 seconds, 1, 2, 5,

10, 15, 20, and 30 minutes, or 1 hour for the measurement period. AVERAGE may be

turned to ON or OFF. When average is on, the temperature and humidity values

displayed are averaged over the measurement period. When average is off, the displayed

values are instantaneous measurements.
RATE T, rate time, determines the time interval over which the measurements are used to

calculate the rate of change, and is selectable between 30 seconds, 1, 2, 5, 10, 15, 20, 30,

45 minutes, or 1 hour. The temperature rate is displayed in the RATE T field of a STATS

display zone. The humidity rate is displayed in the RATE H field of a STATS display

zone. If RATE T is 5 minutes or less, the rate of change is calculated as the average rate

of change using all the measurements obtained during the past RATE T time. If RATE T

is greater than 5 minutes, the rate of change is calculated using the most recent

measurement and the measurement obtained RATE T time ago. If there are as yet

insufficient measurements to calculate the rate of change, only dots are displayed in the

RATE T or RATE H field.