Alarm setting – Fluke 5020A User Manual

Page 61

Advertising
background image

Menu Functions

Alarm Menu

5

5-25

avu034.jpg

Figure 5-33. Alarm

The A D buttons are used to select the menu option. The B C buttons are used to change

the parameter setting. Press the E button to save the new setting. Press the F button

to cancel any changes made and exit to the menu.

Alarm Setting

The ALARM SETTING function is used to enable or disable visible and audible alarm

notifications. The parameters are BEEP and DISPLAY.
When the BEEP parameter is ON, the thermo-hygrometer emits an audible beep during

an alarm condition. For the power alarm it is four short beeps every 20 seconds. For the

battery alarm it is three short beeps every minute.
For the sensor disconnect alarm it is two short beeps every measurement period. For

measurement alarms it is one long beep every measurement period. When the DISPLAY

parameter is ON, an alert message is displayed when an alarm condition occurs.

avu035.jpg

Figure 5-34. Alarm Setting

The A D buttons are used to select the menu option. The B C buttons are used to change

the parameter setting. Press the E button to save the new setting. Press the F button

to cancel any changes and exit to the menu.