Fluke 5020A User Manual

Page 84

Advertising
background image

5020A

Users Manual

6-14

ALARm:BEEP <bool>

Sets the state of the audible alarm enable.
Example: ALAR:BEEP 1
The <bool> parameter turns the alarm beep enable on (1 or ON) or off (0 or OFF). This

command may be password protected (see Password Commands).

ALARm:CLEar

Clears the alarm events.
Example: ALAR:CLE
The alarm clear command clears the alarm events.

ALARm:DATE:FIRSt?

Returns the date of the first alarm event.
Example: ALAR:DATE:FIRS?
Response: 2003,8,10
The response is returned in the format, <year>,<month>,<day>.
If no alarm has been triggered, the response is: 2000,0,0.

ALARm:DATE:LAST?

Returns the date of the last alarm event.
Example: ALAR:DATE:LAST?
Response: 2003,8,10
The response is returned in the format, <year>,<month>,<day>.
If no alarm has been triggered, the response is: 2000,0,0.

ALARm:DISPlay?

Returns the state of the visible alarm enable.
Example: ALAR:DISP?
Response: 0
A value of 1 is returned if the alarm display is enabled. A value of 0 is returned if the

alarm display is disabled.

ALARm:DISPlay <bool>

Sets the state of the visible alarm enable.
Example: ALAR:DISP 1
The <bool> parameter turns the alarm display enable on (1 or ON) or off (0 or OFF).

This command may be password protected (see Password Commands).

ALARm:POWer?

Indicates whether a power loss alarm event has occurred.
Example: ALAR:POW?