Data Menu

5

5-15

Caution

When the memory has been filled, the oldest block of data will

be discarded to allow new data to be recorded.

Measurement storage capacity can range between 28 hours and 2 years depending on the

record period and enabled channels. Refer to Table 5-2, Approximate Data Storage

Capacity.

Table 5-2. Aproximate Data Storage Capacity

Record Period

Storage, 1 channel

Storage, 2 channels

1 second

56 hours

28 hours

10 seconds

20 days

10 days

1 minute

18 weeks

9 weeks

5 minutes

20 months

10 months

10 minutes

24 months

20 months

30 minutes

24 months

24 months

1 hour

24 months

24 months

avu019.jpg

Figure 5-18. Record Setting

Use the A D buttons to select the setting to change. The B C buttons are used to change

the setting parameters. Press the E button to save the new setting. Press the F button

to cancel any changes and exit to the menu.

Data View

The DATA VIEW function allows viewing recorded data in NUMERIC or GRAPHIC

format.