Introduction and Specifications

Specifications

1

1-7

Specifications

Temperature Range

0 °C to 50 °C

Temperature Accuracy

(“H” Model)

16 °C to 24 °C: ±0.125 °C (calibrated)

0 °C to 16 °C, 24 °C to 50 °C: ±0.5 °C (uncalibrated
typical)

Temperature Accuracy

(“S” Model)

15 °C to 35 °C: ±0.25 °C (calibrated)

0 °C to 15 °C, 35 °C to 50 °C: ±0.5 °C (uncalibrated
typical)

Delta Temperature Accuracy

±0.025 °C for ±1 °C changes within 15 °C to 35 °C

Temperature Display Resolution

User selectable up to 0.001 °C (0.0 1°C recorded)

RH Range

0% to 100% RH

RH Accuracy
(“H” Model)

20% to 70% RH: ±1.5% RH (calibrated)

0% to 20% RH, 70% to 100% RH: ±3% RH
(uncalibrated, typical)

RH Accuracy
(“S” Model)

20% to 70% RH: ±2% RH (calibrated)

0% to 20% RH, 70% to 100% RH: ±3% RH
(uncalibrated, typical)

Delta Humidity Accuracy

±1.0% for ±5% changes within 20% to 70% RH

RH Display Resolution

User selectable up to 0.01% (0.1% recorded)

Inputs

Two sensors, each measuring temperature and relative
humidity; each is detachable, cable-extendable, and
interchangeable, with self-contained calibration; each
may be assigned a unique 16-character identification

Display

240 x 128 graphics monochrome LCD, displays
temperature and humidity data graphically, numerically,
and statistically; 16 pre-defined, user-changeable screen
set-ups are included

Memory

400,000 typical individual time-stamped readings
(excluding data card storage)

Alarms

Visual and audio alarms for temperature, temperature
rate, RH, RH rate, and fault conditions

Communications

RS-232 and IR

Data Card Interface

Removable data card for transferring data to a computer;
data can likewise be uploaded from a data card into the
5020A for graphical and statistical display

Enclosure

The 5020A may be wall-mounted (hardware included) or
set on a benchtop

Power

12 V DC from external 100-240 VAC power supply

Battery Backup

Standard 9V battery to allow continued measuring during
power disruptions

Operating Range

0 °C to 50 °C