Measurement commands – Fluke 5020A User Manual

Page 92

Advertising
background image

5020A

Users Manual

6-22

ALARm:TIME:LAST?

Returns the time of the last alarm event.
Example: ALAR:TIME:LAST?
Response: 15,43,19
The response is returned in the format, <hour>,<minute>,<second>.

Measurement Commands

The measurement commands are used to read information on the statistical calculations

and for reading the most recent measurement for a specified channel.

CALCulate:AVERage:CLEar

Resets all running statistics for both channels.
Example: CALC1:AVER:CLE
This command may be password protected (see Password Commands).

CALCulate<chn>:DEWPoint?

Returns the dewpoint measurement on the specified channel in degrees C or F.
Example: CALC1:DEWP?
Response: 3.5
The CALCulate suffix, <chn>, specifies the channel (1 or 2). The value returned is

always in the current temperature units (C or F).

CALCulate<chn>:HINDex?

Returns the heat index calculation for the specified channel in degrees C or F.
Example: CALC1:HIND?
Response: 28.4
The CALCulate suffix, <chn>, specifies the channel (1 or 2). The value returned is

always in the current temperature units (C or F).

CALCulate<chn>:PARameter<num>:AVERage<type>?

Returns the specified statistical calculation for the specified channel.
Example: CALC1:PAR1:AVER4?
Response: 25.186
The CALCulate suffix, <chn>, specifies the channel (1 or 2). The PARamenter suffix,

<num>, specifies temperature (1) or relative humidity (2). The AVERage suffix, <type>,

specifies the statistic: average (1), standard deviation (2), minimum (3), maximum (4),

spread (5), maximum rate (7), or alarm count (8). Temperature values are in the current

temperature unit, C of F.

CALCulate[<chn>]:PARameter[<num>]:AVERage[<type>]:CLEar

Resets all of the running statistics for both channels.
This is the same as the CALCulate:AVERage:CLEar command.