Digital Communications Interface

Commands

6

6-27

DATa:DSTatistics:RHOur? [MIN|MAX|DEF]

Returns the daily statistics reset time.
Example: DAT:DST:RHO?
Response: 12
Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or

default value respectively.

DATa:DSTatistics:RHOur <num>|MIN|MAX|DEF

Sets the daily statistics reset time.
Example: DAT:DST:RHO 0
The <num> parameter specifies the hour for the daily statistics reset time. Specifying

MIN, MAX, or DEF parameter sets the daily statistics reset time to the minimum,

maximum, or default value respectively. This command may be password protected (see

Password Commands).

DATa:DSTatistics:RECord:BTIMe? [<num>]

Returns the reset time for a daily statistics set.
Example: DAT:DST:REC:BTIM?
Response: 0,0,0
The <num> parameter specifies the daily statistics record number. If none is specified

then record 1 is assumed. The response is returned in the format,

<hour>,<minute>,<second>.

DATa:DSTatistics:RECord:CLEar

Clears all daily statistics records.
Example: DAT:DST:REC:CLE
This command may be password protected (see Password Commands).

DATa:DSTatistics:RECord:COUNt? [<MAX>]

Returns the daily statistics record count.
Example: DAT:DST:REC:COUN?
Response: 64
Specifying the MAX parameter returns the maximum number of daily statistics records

than can be held in memory.

DATa:DSTatistics:RECord:DATE? [<num>]

Returns the daily statistics recording date.
Example: DAT:DST:REC:DATE?
Response: 2003,9,12
The <num> parameter specifies the daily statistics record number. If none is specified,

record 1 is assumed. The response is returned in the format, <year>,<month>,<day>.