Blank screen

The screen appears blank when switched on.
Ensure the power adapter is attached properly and the power is switched on.
This problem can also be caused by improper adjustment of the contrast. After the power

is switched on allow the thermo-hygrometer at least one minute to complete its self-test

and power on sequence. Then, try to adjust the contrast using the A D buttons.

Error message at power up

The thermo-hygrometer reports an error during the power up self-test.
The display will remain inoperable during a power outage to preserve battery life.

Measurements will continue to be recorded as long as the battery has a charge.
On power up the thermo-hygrometer performs a series of self-tests to check the operation

of all its components and each of the sensors. The results of the self-tests are reported on

the screen. Normally, all tests report ‘OK’. Below is a table of error messages and

recommended actions to correct the error. Often these errors can be corrected by

completing the actions listed. A failure of one or more component causes error messages

to be reported.
Contact an Authorized Service Center for assistance.

Self-Test Error Messages

Table 8-1. Self-Test Error Messages

Error Message

Recommended Action

SYSTEMS: SERIAL PORT FAILURE

Cycle the power. Contact an Authorized Service
Center.

SYSTEMS: IR PORT FAILURE

Cycle the power. Contact an Authorized Service
Center.

SYSTEMS: DATA CARD FAILURE

Cycle the power. Contact an Authorized Service
Center.

BATTERY: FAILURE

Cycle the power. Contact an Authorized Service
Center.

BATTERY: NOT INSTALLED

Check that a battery is properly installed. Replace
the battery.

CLOCK: INVALID TIME

Set the time/date.

MEMORY DATA: EMPTY

Contact an Authorized Service Center.

MEMORY DATA: QUESTIONABLE

Cycle the power. Contact an Authorized Service
Center.

SENSOR <n>: NOT ATTACHED

Check that the sensors are properly inserted into
their sockets or attached with an undamaged
extension cable.

SENSOR <n>: LOCKED

Attach the correct sensor or, if a new sensor is to be
used, unlock the sensor using CHANNEL-SENSOR
LOCK.