Sensor Lock

The SENSOR LOCK function locks each sensor individually and prohibits the user from

switching sensors on the locked channels. This may be password-protected. The sensor

lock is set to ON or OFF. After selecting ON, a screen is displayed showing the sensor

channel number, serial number, and the message ‘NOW LOCKED’. When the sensor

lock is ON, the thermo-hygrometer measures only on that channel using the sensor that

was attached when sensor lock was set ON. This can be used to prevent inadvertent

attachment of an uncalibrated sensor or one that is not in the desired location. It uses the

sensor serial number to identify the sensor. To exchange a different sensor, the sensor

lock must be set to OFF.

avu007.jpg

Figure 5-6. Sensor Lock

Use the A D buttons to select the sensor to lock. The B C buttons are used to select ON

or OFF. Press the E button to save the setting and exit. Press the F button to cancel

any changes and exit to the menu.

Sensor Cal

The SENSOR CAL function allows the sensor parameters to be edited. This function is

always password-protected. If an incorrect password or no password is entered, the

parameters are displayed but cannot be edited. If a correct password is entered, the

channel, sensor model, sensor serial number, and sensor ID are displayed. Select the

sensor channel of the parameters to edit and then select the E button. The sensor

parameters are displayed for editing.

Caution

The sensor calibration parameters are critical to the accuracy of

the sensor. Do not change them.

The B C buttons are used to select a channel. Press the E button to save the setting and

exit. Press the F button to cancel any changes and exit to the menu.