Fluke 5020A User Manual

Page 129

Advertising
background image

Calibration Procedure

Temperature and Humidity Adjustments

7

7-5

that the measurement errors first be obtained according to the procedure above in section

Temperature and Humidity Measurement Errors. After the Temperature and Humidity

Adjustments procedure is completed, perform the Temperature and Humidity

Measurement Errors procedure again to verify the calibration.

1. Calculate the adjustment

TSL to the temperature slope parameter TSL, and the

new temperature slope parameter:

[

]

(

)

1

3

)

1

(

)

3

(

10

T

T

T

Terror

T

Terror

C

TSL

×

°

=

TSL

TSL

L

TS

+

=

2. Calculate the slope-adjusted temperature errors:

(

)

C

C

Tn

TSL

Tn

Terror

Tn

Terror

°

°

×

+

=

10

25

)

(

)

(

3. Calculate the median slope-adjusted temperature error, the adjustment

TOS to

the temperature offset parameter TOS, and the new temperature offset parameter:

( )

( )

2

2

3

+

=

T

Terror

T

Terror

n

Tmedia

4. Calculate the adjustment

HSL to the humidity slope parameter HSL, and the

new humidity slope parameter:

(

)

(

)

[

]

(

)

1

3

2

3

%

25

RH

RH

RH

RHerror

RH

RHerror

HSL

×

=

HSL

HSL

L

HS

+

=

5. Calculate the slope-adjusted humidity errors:

(

)

(

)

(

)

%

25

%

45

Re

×

+

=

RHn

HSL

RHn

rror

H

RHn

RHerror

6. Calculate the median slope-adjusted humidity error, the adjustment

HOS to the

humidity offset parameter ∆HOS, and the new humidity offset parameter:

(

)

(

)

2

2

3





+

=

RH

THerror

RH

RHerror

n

RHmedia

n

RHmedia

HOS

=

HOS

HOS

S

HO

+

=

7. Enter the new values for TOS (TOS′), TSL (TSL′), HOS (HOS′), and HSL

(HSL′) into the thermo-hygrometer. Also set the DATE to the current date and

set the DUE date for the next calibration. If a power-on warning message is

desired when the DUE date expires, set ALERT to ON. Do not change the serial

number. Press ENTER when prompted to write the new values to the sensor.

Advertising