Starter Kit Instruction Manual

Description

18

4.4.3

Thermo-Electric Cooler (TEC)
and PT100 connections

The pin 1 is recognised by a circle around it and
a thin noon ended circle starts from pin one and
turns around the pins until pin 6.

(1) Pin 1: +Peltier element

(2) Pin 2: - Peltier element

(3) Pin 3: I+ PT100

(4) Pin 4: Sens + PT100

(5) Pin 5: Sens - PT100

(6) Pin 6: I- PT100

Specifications of the Peltier connector

Vacuum case plug reference: . . . . . . . LEMO #
HGG.1b.306.CLLPV.

Cable plug reference: . . . . . . . . . . . . . LEMO #
FGG.1b.306.CLAD72Z.

Plastic protection for the cable: . . .LEMO #
GMA.1b.065.DG.

Cable:

Diameter : . . . . . . . . . . . . . . . . . . . . . 5.8 mm

Section : 6 x 0.25 mm (2 x 0.25 for each
Peltier connections)

or

Diameter : . . . . . . . . . . . . . . . . . . . . . 7.2 mm

Section : 10 x 0.25 mm (2 x 3 x 0.25 for
Peltier connections)

4

5

6

1

2

3

Fig.8: Front view of the Peltier connector