Quantum Instruments Cascade Laser Starter Kit User Manual

Page 50

Advertising
background image

Starter Kit Instruction Manual

Maintenance

49

Oscillator check

[15] Measure the frequency on the pin 7 of
U13 (1).

Note: The value is 28KHz + /- 3KHz

Current/Voltage ratio adjustment

[16] Adjust the trimmer P4 (6) in order to
read B4 = 0.6V (4).

[17] Set the selector (5) to mode Setting +I.

The LCD screen should display 1.00 A.

Note: If it is not the case, adjust it with P2
(2).

[18] Adjust the trimmer P4 in order to read
B4 = 3.0 V.

The LCD screen should display 5.00 A.

Note: If it is not the case, adjust it with P2
(2). Example : 0,6 V ⇒ 0,99 A and 3V ⇒
4,99 A.

[19] Set the switch on the front panel to mode
Setting -I.

[20] Repeat steps [2] and [4] for P5 (7) and
B5 (3).

1

Fig.42: TCU151 main board

2

3

4

Fig.43: TCU151 main board

TCU151

Tem

pera

ture

Con

trolle

r

Alar

m

Res

et

Exte

rnal

Inte

rnal

Rea

l I

Rea

l C

Setti

ng

C

Setti

ng

+I

Setti

ng

−I

+

5

7

6

Fig.44: TC51 front panel

Advertising