Appendix, Chapter 10, 1 bias circut (”bias-t”) – Quantum Instruments Cascade Laser Starter Kit User Manual

Page 56: 1 general, 2 description

Advertising
background image

Chapter 10

Appendix

10.1

Bias Circut (”Bias-T”)

10.1.1

General

The Bias-T allows to apply a constant (DC) cur-
rent to the laser in addition to the pulsed current
(therefore a Bias-T is useless in CW mode). The
current is drawn from the external (user supplied)
power supply through the laser. This current can
be controlled electrically.

Since tuning of a QC laser is done by changing

the temperature of the active zone, the DC Bias cur-
rent can be used to control the emission wavelength
of the laser via its heating effect. The Bias-T there-
fore allows for electrically controlled rapid scanning
of the emission wavelength. The achievable shift is
of the order of 0.1%, with a Bias frequency of up to
100Hz.

10.1.2

Description

The circuit included in the LDD100 pulser unit is
controlled by the twisted black and yellow wires of
the control cable (with the DSUB-9 plug). They
correspond to the shield and center of the IN con-
nector in the former case (positive voltage on yellow
wire). This version has no monitor connection.

Supply

Power

User
DC

+VDC

R

Q

LDD100

L

Bias−T

R

i_b

U_mon

T1

U_in

LLH

Control "IN"

Monitor "MONI"

Low Impedance Line

GND

Fig.53: Bias circuit diagram block

55