Maintenance

51

Current adjustment

[28] Set the selector (3) to mode Setting

C

and adjust the temperature value to 25

C

by means of the Set Temperature 5 tuns
portentiometer P8 (2).

[29] On the external power supply, rise slowly
the voltage to 30 V and the current to 1.5 A.

→ The tension should stabilize at 30 V

Note: If the current is too high, stop the test
and verify the transistors Q1 to Q4 (1) and
their associated resistors.

There might be a soldering problem. In this
case, fix it and repeat the procedure from step
[10].

[30] Set the selector (3) to mode Real I.

→ Value displayed: ≤ 1A.

[31] Set the selector (3) to mode Real

C.

→ Value displayed should tend to the defined

value.

[32] Wait for a certain time and check the
Peltier temperature with a thermometer

[33] Set the selector (3) to mode Setting +I.

[34] Increase the limit of positive current to
3.5A by means of the +I trimmer (4) located
on the front panel.

[35] Increase the limit of negative current to
5A by means of the -I trimmer (5) located on
the front panel.

[36] Change the temperature reference to -25

C and check the displayed current.