Starter Kit Instruction Manual

Maintenance

38

8.2

Replacement procedures

8.2.1

Quantum Cascade Laser replace-
ment

Before beginning

[1] Switch the QCL Starter Kit OFF

Material needed

• 1 set of Allen keys.

• 1 ohmmeter.

• 1 QCL laser.

• a pair of tweezers.

Procedure

To replace the Quantum Cascade Laser, proceed as
follows:

[1] Remove the top cover (2) of the Laboratory
Laser Housing (LLH100) by unscrewing the
screws (1)

[2] Remove the PET U-shaped holder (5) by
unscrewing the fixing screws (3)

CAUTION ! Pay attention NOT to drop the
laser!

[3] Remove the laser (4). Hold the laser on
the gilded ceramic pads with the tweezers pins
along the axis of the largest length of the cop-
per submount

! WARNING ! Take especially care not
to touch the active region of the laser
or the wire bonds!

2

3

4

5

1

Fig.26: Removing the laser