Starter Kit Instruction Manual

Maintenance

48

Temperature Reference offset adjustment

[10] Adjust B13 (3) to 0.000 with P3 (4)

[11] Set the temperature with the Set Tem-
perature potentiometer (Fig.40 (7)) to +5

C

→ The temperature reference signal B13 should

be ≥4.3V

[12] Set the temperature with the Set Tem-
perature potentiometer (Fig.40 (7)) to -5

C

→ The temperature reference signal B3 should

be ≤4.3V

[13] Check the switch range by setting the
Temperature Reference at + 2

C and -2

C

B7 = -5V and +5V

B6 = +5V and -5V

[14] Check the output voltage of the transis-
tors

→ The signals on R78 and R81 should tog-

gle simultaneously with the inversal signals on
R79 and R80

CAUTION ! The signals B6 and B7 (2)
must never be at 0V simultaneously. Other-
wise there is a risk of short-circuit!

1

2

3

4

Fig.41: TCU151 main board