Starter Kit Instruction Manual

Maintenance

52

Miscellaneous functions testing Temperature
limit testing

[37] Decrease the temperature limit threshold
from 0.7 V to 0.3 V.

Note: For more details, see Temperature limit
threshold on page 50.

[38] Increase the temperature reference to 35

C.

→ When the threshold is overpassed, the red

LED (1) in front panel should lit.

[39] Wait the real temperature drops 10

C

below the threshold.

[40] Reset the instrument with the RESET
button (2).

→ The red LED (1) should turn off.

[41] Reset the temperature limit threshold to
0.7 V.

Note: For more details, see Temperature limit
threshold on page 50.

Monitoring

[42] Check the output voltage on the rear in-
strument BNC connector (5).

Note: For 35

C, the voltage should be 350

mV.

External reference

[43] Plug an external power supply to the Ex-
ternal Reference BNC connector (6).

[44] Set the voltage to 3.5 V.

[45] Set the switch External/Internal (3)
to External.

[46] Set the selector (4) to mode Setting

C.

→ The temperature displayed (2) should be

035

C.

[47] Set the selector (4) to mode Real

C.

→ The temperature displayed (2) should fol-

low the reference temperature.

TCU151

Tem

pera

ture

Con

trolle

r

Alar

m

Res

et

Exte

rnal

Inte

rnal

Rea

l I

Rea

l C

Setti

ng

C

Setti

ng

+I

Setti

ng

−I

+

1

2

3

4

Fig.49: TCU151 front panel

0

1

US

E O

NLY

WIT

H

250

V

FU

SE

S / E

MP

LOY

ER

UN

IQU

EM

EN

T A

VEC

DE

S F

US

IBLE

S D

E 2

50v

110

−12

0V

22

0−2

40

V

UTP

UT

Max

: 15

V/6A

MO

NIT

OR

ING

10

mV

/ C

Zou

t: 20

0 oh

m

Exte

rna

l Re

f.

100

mV

/ C

Zin:

1 M

ohm

Inte

rloc

king

6

5

Fig.50:TCU151 rear panel

Advertising