5 qcl pulser switching unit(ldd100), 1 description, 2 specifications – Quantum Instruments Cascade Laser Starter Kit User Manual

Page 21

Advertising
background image

Starter Kit Instruction Manual

Description

20

4.5

QCL pulser switching unit(LDD100)

4.5.1

Description

The switching unit is based on dedicated power
MOS-FETS with up to 30A current compatibilities
and 60V. In addition, the circuit is limited in power
dissipation by the cooling of the unit. During op-
eration, the temperature of the case should not ex-
ceed 60C. While powering usual QC diodes, these
limitations are never reached because the laser has
usually much lower power ratings.

! WARNING !
An important feature of this unit is that

both lines going to the laser are ”HOT”, i.e.
have a postive voltage respective to the case.
this feature should be kept in mind when de-
signing the laser holder. This one should be
insulated and have a low capacitance(< 100pF)
towards the ground.

The QCL pulse switching unit is composed of

the following connectors:

(1) Low impedance connector

(2) Drive cable connector

(3) Monitoring BNC connector

4.5.2

Specifications

Voltage : . . . . . . . . . . . . . . . . . . . . . . max 60 VDC

Current : . . . . . . . . . . . . . . . . . . . . . . max 30 ADC

Repetition rate : . . . . . upto 1MHz, limited by
average current

Rising/falling edge: . . typical 8ns (max 12ns)

Minimum pulse width: . . . . 16ns, typical 20ns

Maximum pulse width: 1µs, limited by power
supply

Support power: . . . . . . . . . . . . given by TPG128

Trigger IN: . . . . . . . . . . . . . . . . . . . . . . . . TTL 50Ω

Propagation delay: . . . . . . typical 35ns <40 ns

Size: . . . . . . . . . . . . . . . . . . . . . 90 x 130 x 50 mm

3

! WARNING !
The connection between the laser and the

QCL pulse switching unit is floating. It must
not be connected with the ground.

1

2

3

connected to + hi High Voltage

Pulse on top, Bottom

Output

Current control Max 60V

Power supply 12V

Input

LDD 100

Monitor 20 A/V, Z50 Ohm

Fig.9: QCL pulser switching unit front and rear view