Installation

27

5.4

Setting the appropriate AC
voltage on TCU151

The TCU151 temperature controller works on 110/120V
or 220/240V if the fuse holder is oriented in the ap-
propriate position.

Note: There are two rates:

110-120V, 60Hz (USA)

220-240V, 50Hz (Europe)

5.4.1

Procedure

To set the appropriate voltage on the TCU151 tem-
perature controller, proceed as follows:

[1] Pull out the fuse holder (1) from the socket.

[2] Verify the orientation of the fuse holder to
get the appropriate voltage. The arrow on the
fuse holder (3) should correspond to the mark
on the socket (2).

[3] Insert the fuse holder (1) into the socket.