Starter Kit Instruction Manual

Description

21

4.6

QCL pulse switching measuring unit

4.6.1

Generalities

The measuring circuit contained in the LDD100
provides information about various laser pulse pa-
rameters (peak voltages, supply voltages, duty cy-
cle, frequency). They are only estimated values,
since exact measurement of short and strong pulses
with diodes and averaging circuitry is difficult.

Keep in mind that you should always measure

the voltages on the LLH100 if you need accurate
time and voltage data (see final paragraph). How-
ever, these data are useful for monitoring and surveil-
lance purposes, and to give a rough estimation of
the current parameters.

4.6.2

Measured voltages

• UHV: 1/2 of the average voltage, respective

to VHT (user supplied high voltage)

• ULH: 1/2 of the average laser anode voltage,

respective to VHT

• ULL: duty cycle dependent peak laser cathode

voltage, respective to VHT

• UD: duty cycle dependent peak transistor drain

voltage, respective to VHT

• UPI: 1/2 average internally reshaped drive volt-

age, respective to ground

• UPT: 1/2 average of a 37ns fixed length pulse,

respective to ground

4.6.3

Timing data

Pulse frequency calculation

UPT can be used to calculate the pulse frequency
as follows:

f = f

0

2

U P T −ν

P low

ν

P high

−ν

P low

Where f

0

is a frequency constant, v

P low

and

v

P high

are the TTL pulser voltage limits.

Actual values: f

0

=

1

37

ns , v

P low

=0.01V, v

P high

=5V