8.3

TCU151 temperature con-
troller interlock setting

8.3.1

Generalities

The interlock is located on the main board.

Note: By default, the interlock is set to NC (Nor-

mally Close) contact.

8.3.2

Before beggining interlock set-
ting

[1] Switch the instrument OFF.

[2] Unplug the power cord.

[3] Remove the cover (1) to acces the main
board (2).

8.3.3

Needed material

• 1 set of screwdriver.

• 1 tweezer.

8.3.4

Procedure interlock setting

To set the type of interlock, proceed as follows:

[1] Locate the J9 Interlock connector.

[2] Place the associated jumper according to
the type of desired interlock.

Note: The common pin is on the center (4)
and by placing the jumper on a side defines
the interlock type as:

– for a NC contact: jumper on pins (4) and

(5)

– for a NO contact: jumper on pins (3)

and (4)