Starter Kit Instruction Manual

Maintenance

47

Sensing level and range adjustement

[4] Replace the cable connected to J8 (4) with
the 100Ω 1% reference resistor in the follow-
ing way:

Connect one side of the resistors to pins 1 and
2, the other side to pins 3 and 4.

[5] Measure the voltage on Zn3/R10 (3).

Note: The value should be:

- V = 1.22 to 1.25V.

[6] Measure the voltage at the intersection of
R19/R35 (5).

Note: The value should be: V=100mV (ad-
justable with the potentiometer P1 (2)).

Note: If the range is too small, replace the
resistor R1 (1) = 1.3kΩ with 1.4kΩ.

[7] Set the selector (9) to display Real

C.

[8] Adjust the trimmer P6 (6) in order to ob-
tain the value of 000

C on the screen (8).

[9] Vary the Temperature Reference by using
the Set Temperature 5 turns potentiometer
(7) located on the front panel.

Note: The range displayed should be -074 to
+ 074.