Starter Kit Instruction Manual

Maintenance

50

CAUTION ! For the further steps of this

check, limit the positive current to 1 A and
the negative one with 1.2 A by means of the
5 turns potentiometer)

Temperature limit threshold

[21] Measure the voltage on pin 2 of U10
(1).

Note: The value should be: 0.7V (= 70

C)

(If needed, adjust it with the trimmer P9(2)).

[22] Power off the instrument.

Starting the power section of the instru-
ment.

[23] Remove the 100 E reference resistor from
J8 (4) and plug the output cable J8.

[24] Plug the external cable into the LLH100,
and verify that the Pt100 is proprely mounted
in the latter.

[25] Use an external power supply (0-30VDC/3A)
to simulate the internal power supply.

[26] Limit the current to 200 mA and connect
it to main board J1 connector (3).

[27] Power on the instrument and the external
power supply.

1

2

Fig.45: TCU151 main board

3

4

Fig.46: TCU151 main board