Starter Kit Instruction Manual

Description

25

4.8

QCL temperature controller(TCU151)

4.8.1

Description

The TCU151 is used to control the laser’s temper-
ature inside the Laboratory Laser Housing.

It uses a PT100 sensor to measure the tempera-

ture of the cold plate and maintains a pre-set tem-
perature either from a front panel knob or from a
user supplied voltage.

4.8.2

Specifications

Voltage: . . . . . . . . . . . . . . . . . . . . . . . 120-240V AC

Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . 50-60 Hz

Fuse:

. . . . . . . . . . . . . . . . . . . . 1.6 A T (220-240V)

. . . . . . . . . . . . . . . . . . . 3.2 A T (110 -120V)

Peltier current maximum: . . . . . . . . . . . . . . . . 5A

Temperature control range: . -65

C to 65

C

Temperature measurement: . . . PT100 4 wires

Temperature monitor: . . . . 10mV/

C, Zout =

200Ω

In addition, allow for 15 min. warm-up of
TCU151 to get stable readings

External driving temperature: . -6.5V to 6.5V
(-65

C to 65

C), Zin =1MΩ

4.8.3

Peltier and PT100 connections

The rear connector pinout is listed according to the
following items:

(1) Pin 1: +Peltier element

(2) Pin 2: - Peltier element

(3) Pin 3: +I PT100

(4) Pin 4: +Sense PT100

(5) Pin 5: -Sense PT100

(6) Pin 6: -I PT100

(7) Pin GND: Ground

TCU151

Tem

pera

ture

Con

trolle

r

Alarm

Res

et

Exte

rnal

Inter

nal

Rea

l I

Rea

l C

Settin

g

C

Settin

g +I

Settin

g −I

+

Fig.11: TCU151 temperature controller

1

2

1

7

3

4

5

6

Fig.12: TCU151 Peltier Connector

Advertising