Starter Kit Instruction Manual

Maintenance

45

[12] Connect oscilloscope on the output Out1.
Load with a 50Ω . Choose a repetition period
of about 5 µs.

[13] Pot 10 turns Duration on 10. Adjust
capacitor Max duration (2) to have a output
pulse of 200 ns +/-5 ns.

[14] Pot 10 turns Duration on 0. Adjust the
pot Out 0ns adj (1) in order that the output
pulse is 0ns. The pulse is 0 ns long when its
shape is triangular with an 1,1 V amplitude
(half of the maximum value).

[15] Repeat points [13] and [14] until the val-
ues are OK.

[16] Check the output Out 2, check the oper-
ation of Gate in. The input Gate in open
(high TTL level) = gate open, the output sig-
nal is present. Input Gate in short-circuited
(low TTL level) = gate closed, the output
signal is zero.

[17] Check the +12 V is present on Lemo 00
(+12 V on the centre connector).

2

1

Fig.36: TPG128 main board