Chapter 7

Troubleshooting

7.1

Chapter overview

This chapter sets out troubleshooting instructions
for ensuring safe and trouble-free operation of the
Quantum Cascade Laser Starter Kit system.

7.2

Principle

The personnel must have read and understood this
documentation before carrying out any activity what-
soever with the Quantum Cascade Laser Starter
Kit. In case of unclear information, please contact
the distributor.

The power supply reaches max cur-

rent for a very low voltage

- The low impedance line or the QCL is

shorted to ground.

Check the cabling.

Replace the laser if necessary.

- The DC power supply is NOT floating.

Change the cabling in order to leave

the laser floating.

- The laser tightening device is too tight-

ened and the QCL laser is shorted to
ground.

Loose the laser fixture device.