background image

Starter Kit Instruction Manual

Maintenance

43

8.4

Calibaration procedures

8.4.1

TPG128 calibration

Generalities

The present section describes procedure to be car-
ried out for particular calibration of the TPG128.

Before beginning

[1] Switch the instrument OFF.

[2] Remove the cover (1).

[3] Switch the instrument ON.

! WARNING ! Keep in mind that dur-

ing the calibration, the instrument is alive
(powered on)

Needed material

• 1 set of screwdriver.

• 1 oscilloscope.

• 1 voltmeter.

Procedure calibration

The generator should be calibrated as followed:

[1] Verify the +5 V supply.

[2] During the first test, set all potentiometers
and adjustable capacitors to their medium val-
ues.

[3] Adjust to the potentiometer 50 ns adj
(3) in order to obtain a 500 ns long pulse on
TP2 (2).

[4] Connect an oscilloscope allowing the mea-
surement of the signal’s period on the output
Trig out.

TPG

128

−TT

L P

ulse

Gen

era

tor

OF

F

ON

Pow

er

0.2

to

2.2

m

s

0.5

to

10.5

m

s

5 to

105

m

s

0 to

200

ns

−−−

− −

−−−

−Pe

riod

−−−

−−−

−−−

−Du

ratio

n −

Gate

IN

Trig

OU

T

Out

1

Out

2

50

ohm

s

50

ohm

s

1

50

ohm

s

ma

x 6

0m

A

+12

V

0 V

Fig.33: Removing the top cover

2

3

Fig.34: TPG128 main board