10 rs-232 configuration, 11 gpib address, 12 gpib connections – B&K Precision 4034 - Manual User Manual

Page 14

Advertising
background image


14

DB-9 pin Name

Note

1
2
3
4
5
6
7
8
9

-

TXD
RXD

-

GND

-

RTS
CRS

-

-

Transmit Data

Receive Data

-

Signal ground

-

Request to Send

Clear to send

-

*Note: Use a Null-modem or cross over cable (pin 2 and 3 switched) in order to communicate with instrument.

2.10 RS-232 Configuration


The instrument use 8 data bits, 1 stop bit, no parity and baud rate selectable from 2400 to 115K (2400, 4800,
9600, 19200, 38400, 57600, 115200). By default, the instrument is set at 19200-8-N-1.

Note: If 115K baudrate speed is used, ensure that the RS232 cable is short and can support this speed. Otherwise,

there may be some instability and intermittent data transmission failure between the interfacing computer and
the instrument.

2.11 GPIB Address

The address can be changed from the front panel by using the "UTILITY" menu.


2.12 GPIB Connections

The rear panel GPIB connector is an AMPHENOL 57-10240 or equivalent, and connects to a standard IEEE-488
bus cable connector. The GPIB line screens are not isolated from chassis and signal ground.