Specifications – B&K Precision 4034 - Manual User Manual

Page 7

Advertising
background image


7

Specifications

NOTE


Specifications listed in manual are applicable after a powered 30 minute warm-up into a 50

Ω load

All timing characteristics are measured at 50% of amplitude with fastest edge
Specifications are verified according to the performance check procedures.
Specifications not verified in the manual are either explanatory notes or general performance characteristics only.
Specifications and information is subject to change without notice. For the most current and correct data please
visit

www.bkprecision.com


MODELS

4033

4034

CHANNELS

1

2

FREQUENCY

0.1 Hz to 50 MHz

TIMING CHARACTERISTICS
PERIOD

Range (single pulse)

20 ns to 10 s (50 MHz to 0.1 Hz repetition rate)

Range (double
pulse)

40 ns to 10 s (25 MHz to 0.1 Hz repetition rate)

Resolution

Up to 6 digits, limited to 10 ps

Accuracy

± 0.01 %

Jitter

< 0.01 % of setting +20 ps on Period, Width and Delay

WIDTH

Range

10 ns to (Period – 10 ns)

Resolution

Up to 6 digits, limited to 100 ps

Accuracy

±(0.5% of setting +500 ps)

Double Pulse

±(0.5% of setting +3 ns) for the second pulse

DELAY

Range

0ns to (Period – Width – 10 ns)

Resolution

Up to 6 digits, limited to 100 ps

Accuracy

±(0.5% of setting +500 ps)

DUTY
CYCLE

Range

1 to 99%

Resolution

3 digits (0.1%)

Accuracy

Limited by width and pulse accuracy

OUTPUT CHARACTERISTICS
AMPLITUDE

High Level
Range

-9.90 V to +10 V into 50 ohms load (-19.80 V to +20 V into open circuit)

Low Level
Range

-10 V to +9.90 V into 50 ohms load (-20 V to +19.80 V into open circuit)

Amplitude Range 0.1V to 10V p-p into 50 ohms load (20 Vp-p max into open circuit)
Resolution

3 digits limited to 10 mV

Accuracy

± 1% of setting ± 10 mV into 50 ohms

Aberrations

<5% + 20 mV into 50 ohms load, for pulse levels between

±5 V

Output
Resistance

50 ohms

Offset Accuracy

± 1% ± 25 mV

OPERATING MODES

Continuous Output continuous at programmed period rate

Triggered

Output quiescent until triggered by an internal, external, GPIB or manual
trigger, then generates one cycle at programmed period rate

Gated

Same as triggered mode except pulses are output for the duration of the
gated signal. The last cycle started is completed

Burst

Same as triggered mode for programmed number of cycles from 2 to
999,999 as set by the N-BURST function

External Width Trigger duration and rate sets pulse width and repetition