4.12.1.6 Predefined Voltage Levels

This command is used to set the pulse voltage levels to predefined values. Four predefined values are
available as follows:

CMOS: High level 5V; Low level 0V
TTL:

High level 2.4V; Low level 0.4V

ECL:

High level –0.8V; Low level –1.8V

USER: User-defined levels, as set using the PHIGH and PLOW commands

COMMAND TYPE:

Setting only

SETTING

Syntax:

[:SOURce]:VOLTage[:LEVel][:IMMediate]:PREDefined<ws><option>

Arguments:

Type:

Character

Options:

CMOS, TTL, ECL, USER

Examples:

:VOLT:PRED ECL

4.12.1.7 High Voltage Limit

This command is used to set the high limit of the pulse.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:VOLTage:LIMit:HIGH<ws><high limit>[units]

Arguments:

Type:

NRf

Units:

MV, V (default)

Range:

-9.5V to +10V

Rounding:

To 10mV

Examples:

:VOLT:LIM:HIGH 4V

QUERY

Syntax:

[:SOURce]:VOLTage:LIMit:HIGH?

Examples:

:VOLT:LIM:HIGH?

Response:

NRf

CONSIDERATIONS:

The high limit cannot be set to less than the high level.

4.12.1.8 Low Voltage Limit

This command is used to set the low limit of the pulse.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:VOLTage:LIMit:LOW<ws><low limit>[units]

Arguments:

Type:

NRf

Units:

MV, V (default)

Range:

–10V to +9.5V

Rounding:

To 10mV

Examples:

VOLT:LIM:LOW 4V