15 pulse definitions – B&K Precision 4034 - Manual User Manual

Page 33

Advertising
background image


33

Delay max

= [(Period * 0.99) - Width –10 ns], but not > 9.80000 s

Delay min

= (Width + 10 ns)

Double Pulse Transition Time Restrictions

Width must be > 1.3 * Leading Edge

(Delay - Width) must be > 1.3 * Trailing Edge

[Period - (Delay + Width)] must be > (1.3 * Trailing Edge)


Internal Trigger Burst Mode


(0.99 * Trig Rate) must be > (Period * Burst Count)



3.15 Pulse Definitions


Pulse Definitions – High and Low Levels

- -

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

- -

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

- -

-

-

-

-

- -

-

-

-

-

- - - - - - - - - - - - - - - - -

TRIG OUTPUT

50%

HIGH LEVEL

LOW LEVEL

AMPLITUDE 50%

90%

10%

50%

90%

10%

WIDTH

3

rd

Corner

1

st

Corner

LEADING
EDGE

TRAILING
EDGE

Advertising