2 high voltage level, 3 low voltage level – B&K Precision 4034 - Manual User Manual

Page 50

Advertising
background image


50

Range:

0.1Hz to 50MHz

Rounding:

To the resolution of the range.

Examples:

:FREQ 5KHZ

:FREQ 5E3

QUERY

Syntax:

[:SOURce]:FREQuency[:CW|:FIXed]?

Examples:

:FREQ?

Response:

NR3

CONSIDERATIONS: FIXed is an alias for CW.

4.12.1.2 High Voltage Level

This command is used to set the high level of the pulse.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:VOLTage[:LEVel][:IMMediate]:HIGH<ws><high level>[units]

Arguments:

Type:

NRf

Units:

MV, V (default)

Range:

-9.5V to +10V

Rounding:

To 10mV

Examples:

VOLT:HIGH 4V

QUERY

Syntax:

[:SOURce]:VOLTage[:LEVel][:IMMediate]:HIGH?

Examples:

:VOLT:HIGH?

Response:

NRf

CONSIDERATIONS:

1)

The high level must be greater than the low level.

2)

The difference between the levels must conform to 0.5V

≤ difference ≤ 10V

3)

The high level may not exceed the high limit.

4.12.1.3 Low Voltage Level

This command is used to set the low level of the pulse.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:VOLTage[:LEVel][:IMMediate]:LOW<ws><low level>[units]

Arguments:

Type:

NRf

Units:

MV, V (default)

Range:

-10V to +9.5V

Rounding:

To 10mV

Examples:

:VOLT:LOW 4V

QUERY

Syntax:

[:SOURce]:VOLTage:[:LEVel][:IMMediate]:LOW?

Advertising