4.10.5.1 Command Errors

A command error is in the range -199 to -100, and indicates that a syntax error was detected. This
includes the case of an unrecognized header. The occurrence of a command error causes the CME bit
(bit 5) of the Standard Event Status Register to be set.

Code

Description

-100

Command Error

-101

Invalid character

-102

Syntax error

-103

Invalid separator

-104

Data type error

-105

GET not allowed

-108

Parameter not allowed – More parameters than allowed were received

-109

Missing parameter – Fewer parameters than necessary were received

-110

Command header error

-111

Header separator error

-112

Program mnemonic too long – The mnemonic must contain no less than 12 characters

-113

Undefined header

-114

Header suffix out of range

-120

Numeric data error

-121

Invalid character in number

-123

Exponent too large – IEEE 488.2 specifies maximum of 32000

-124

Too many digits – IEEE 488.2 specifies maximum of 255 digits in mantissa.

-128

Numeric data not allowed – A different data type was expected

-131

Invalid suffix

-134

Suffix too long – A maximum of 12 characters are allowed in a suffix

-138

Suffix not allowed

-140

Character data error

-141

Invalid character data – Incorrect character data were received

-144

Character data too long – Character data may contain no more than 12 characters

-148

Character data not allowed

-158

String data not allowed

-168

Block data not allowed

-178

Expression data not allowed


4.10.5.2 Execution Errors

An execution error indicates that the device could not execute a syntactically correct command, either
since the data were out of the instrument's range, or due to a device condition. The EXE bit (bit 4) of
the Standard Event Status Register is set on occurrence of an execution error.

Code

Description

-200

Execution error

-201

Invalid while in local – An attempt was made to change an instrument setting while the
instrument was in the LOCAL state

-211

Trigger ignored – The GET or *TRG common command was ignored due to the device
not being in the correct state to execute the trigger

-221

Settings conflict – The parameter is out of range due to the current instrument state

Advertising