QUERY

Syntax:

[:SOURce]:VOLTage:LIMit:LOW?

Examples:

:VOLT:LIM:LOW?

Response:

NRf

CONSIDERATIONS:

The low limit cannot be set greater than the low level.

4.12.1.9 Pulse Period

This command is used to set or query the period of the pulse.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:PULSe:PERiod<ws><period>[units]

Arguments:

Type:

NRf

Units:

S (seconds), MS (milliseconds), US (microseconds), NS nanoseconds

Range:

20NS to 10S

Rounding:

To current resolution

Examples:

:PULS:PER 1US

:PULS:PER 400E-6

QUERY

Syntax:

[:SOURce]:PULSe:PERiod?

Examples:

:PULS:PER?

Response:

NRf

CONSIDERATIONS:

The allowed range of the period will be determined by the values of the width, delay, and
transition times.

4.12.1.10 Pulse Width

This command is used to set or query the value of the pulse width. If the duty cycle is ON when the
width command is sent, it is then set to OFF, and changes in the period will no longer affect the width.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:PULSe:WIDTh<ws><width>[units]

Arguments:

Type:

NRf

Units:

S (seconds), MS (milliseconds), US (microseconds), NS nanoseconds

Range:

10NS to 9.89999S

Rounding:

To current resolution

Examples:

:PULS:WIDT 25NS

:PULS:WIDT 200E-9

QUERY

Syntax:

[:SOURce]:PULSe:WIDTh?