11 ieee 488.2 common commands and queries, 1 system data commands, 2 internal operation commands – B&K Precision 4034 - Manual User Manual

Page 45: 1 *rst - reset command, 2 *tst? - self-test query, 3 synchronization commands, 1 *opc - operation complete command

Advertising
background image


45




For Model 4034

500 Trigger rate short on channel 1
501 Trigger rate short on channel 2
510 Output overload on channel 1
511 Output overload on channel 2

"Trigger rate short" means that the period of the waveform is larger than the value of the internal trigger rate.
Thus not every trigger will generate a cycle (or burst) of the waveform.


4.11 IEEE 488.2 Common Commands and Queries

4.11.1 System Data Commands

The identification query command, *IDN?, enables unique identification of the device over the GPIB.
It returns a string with four fields:

Manufacturer name
Model name
Serial number (0 if not relevant)
Version number

4.11.2 Internal Operation Commands

4.11.2.1 *RST - Reset Command

The Reset command resets the device and returns it to the factory default power-up state.

Command Type: Common Command

Syntax:

*RST

4.11.2.2 *TST? - Self-Test Query

The self-test query causes an internal self-test to be performed. This test consists of checking the status
of the period, pulse and output cards.

Command Type Common Query

Syntax

*TST?

Response

ASCII 0 if test passes
ASCII 1 if test fails

4.11.3 Synchronization Commands

4.11.3.1 *OPC - Operation Complete Command

The operation complete command causes the device to generate the operation complete message in the
Standard Event Status Register, on completion of the selected device operation.