3.10 Power-On Settings

At power-on, the pulse generator performs a diagnostic self-test procedure to check itself for errors. When the pulse
generator finishes the diagnostic self-test routine, it enters the local state (LOGS) and assumes power-on default
settings if the POWER-ON setting is at 0. You can program the pulse generator for any settings you want at power
on, as described earlier in this section.

The factory default settings are:

Power-on Default Settings

Key Functions

Values

Comments

PERIOD

500 ns

Pulse Period

WIDTH

200 ns

Pulse Width

DELAY

0 ns

Pulse delay from Sync out

DPDELAY

5 us

Delay between pulses in double

pulse mode

HILVL

2.5 V

Pulse high level

LOLVL

-2.5 V

Pulse low level

MODE

CONT

Pulse mode

N-BURST

2

Waves per burst

SLOPE

POS

Positive external trigger slope

TLVL

1 V

External trigger level

TRIG SOURCE

MAN

Trigger source

INT TRG RATE

1 ms

Internal trigger rate

OUPTUT

OFF

Output disabled

PULSE MODE

Normal

Normal single pulse
output

MODULATION

OFF

Modulation execution

RISE

5 ns

Pulse rise time

FALL

5 ns

Pulse fall time

Table 3-2

NOTE: Power-on settings cannot restore the status of output at power-on, meaning if the output is ON, power-on settings
cannot recall it to be ON at start up. This setting will always remain OFF and power on, which is same as the default setup
indicated above in Table 3-2. Although the output status can be stored into memory for recall using the store/recall
functions, it cannot be recalled for a power-on setting start up. This is due to safety concerns as sensitive devices that are
connected to the outputs of the generator may accidentally be damaged at power-on if the power-on configurations are not
set properly (i.e. Amplitude level set too high for power-on may easily damage a sensitive device by accident).



3.11 Memory


The pulse generator uses a non-volatile FLASH memory for storing front panel settings. Up to 100 front panel
settings can be stored.

3.12 Displaying Errors

At power-on, the waveform generator performs a diagnostic routine to check itself for problems. If the diagnostic